. I-OPA for PHAROS.

Optical Parametric Amplifier I-OPA, Industrial grade

img

Features

  • Manually tunable wavelength
  • Mounts directly on laser head and integrated into the system
  • Bandwidth limited or short‑pulse configurations available
  • CEP option

Overview

I‑OPA is an optical parametric amplifier of white‑light continuum pumped by the PHAROS laser. This OPA is focused on generating long‑term stable output with reliable and hands‑off operation. Manually tunable output wavelength extends the application possibi­lities of a single laser source instead of requiring multiple lasers based on different technologies.

In comparison to standard ORPHEUS line of devices, the I‑OPA lacks only computer controlled wavelength selection. On the other hand, in‑laser mounted de­sign provides mechanical stability and eliminates the effects of air‑turbulence, ensuring stable long‑term performance and minimizing energy fluctuations.

Output Characteristics
MODEL I‑OPA I‑OPA‑F I‑OPA‑ONE I‑OPA‑CEP
Based on OPA ORPHEUS ORPHEUS‑F ORPHEUS‑ONE
Pump pulse energy 10 – 500 µJ 10 – 400 µJ 20 – 500 µJ 150 – 500 µJ
Pulse repetition rate Up to 1 MHz Up to 100 kHz
Signal 630 – 1030 nm 650 – 900 nm 1350 – 2060 nm
Idler 1030 – 2600 nm 1200 – 2500 nm 2060 – 4500 nm 1400 – 2500 nm
Conversion efficiency signal+idler combined > 12 % > 12 % > 15 % > 10 %
Pulse energy stability < 2 % STD over 1 min. 1) 650 – 950 nm
1150 – 2000 nm
650 – 850 nm
1350 – 2000 nm
1500 – 3500 nm 1400 – 2000 nm
Pulse bandwidth 3) 100 – 150 cm-1 200 – 600 cm-1 80 – 200 cm-1 ~150 cm-1
Pulse duration 1) 150 – 250 fs 30 – 80 fs 200 – 300 fs < 200 fs
Applications Micro-machining
Microscopy
Spectroscopy
Nonlinear microscopy
Ultrafast spectroscopy
Micro-machining
Mid-IR generation
OPCPA front-end
  1. 1)Better stability can be specified for a specific wavelength (e.g.<1 % STD at 800 nm).
  2. 2)Output pulse duration depends on wavelength and pump laser pulse duration.
  3. 3)I‑OPA‑F outputs broad bandwidth pulses which are compressed externally.
Comparison with other Femtosecond and Picosecond Lasers